吴佳涛,陈国茜,许学猛,徐 鹉,关颖欣,吴祖贵.MyotonPRO用于评估健康青年人股四头肌肌肉弹性的信度研究[J].中国康复医学杂志,2021,(1):57~62
MyotonPRO用于评估健康青年人股四头肌肌肉弹性的信度研究    点此下载全文
吴佳涛  陈国茜  许学猛  徐 鹉  关颖欣  吴祖贵
广州中医药大学,广东省广州市,510006
基金项目:广东省省级工业与信息化发展专项资金(粤财工〔2016〕69 号);许学猛广东省名中医传承工作室建设项目(粤中医办函[2017]17);膝骨关节病中医药优势病种研究项目(粤中医办函[2018]6号)
DOI:10.3969/j.issn.1001-1242.2021.01.010
摘要点击次数: 1034
全文下载次数: 703
摘要:
      摘要 目的:检验MyotonPRO肌肉弹性测试仪测量健康青年人股直肌、股内侧肌及股外侧肌肌肉弹性的复测信度和测试者间信度。 方法:本研究纳入了38例健康青年人,采用MyotonPRO评估放松状态下的股直肌、股内侧肌及股外侧肌肌肉弹性,计算组内相关系数(ICC)、绝对信度,两次测量结果绘制成Bland-Altman图并进行系统偏倚分析。 结果:复测信度:组内相关系数值介于0.76—0.90,测量标准误(SEM)值介于0.05—0.09,SEM%介于4.63%—7.11%,最小真正改变量(MDC)值介于0.15—0.25,MDC%介于12.83%—19.70%;测试者间信度:ICC值介于0.86—0.97,SEM值介于0.03—0.08之间,SEM%介于3.42—5.84,MDC介于0.09—0.23,MDC%介于9.48%—16.19%。Bland-Altman图形分析显示无系统性误差。 结论:MyotonPRO用于健康青年人股直肌、股内侧肌、股外侧肌肌肉弹性测量具有良好的信度,可为临床提供客观依据。
关键词:MyotonPRO  股四头肌  肌肉弹性  信度
Measurement reliability of MyotonPRO for elasticity of quadriceps femoris in healthy youth    Download Fulltext
Guangzhou University of Chinese Medicine,Guangdong,Guangzhou,510006
Fund Project:
Abstract:
      Abstract: Objective:To examine the intra-rater reliability and inter-rater reliability of MyotonPRO measurement of the muscle elasticity of rectus femoris, vastus medialis and vastus lateralis in healthy young adults. Method:Thirty-eight healthy subjects were recruited and the muscle elasticity (logarithmic decrement) of rectus femoris, vastus medialis and vastus lateralis in a relaxed state was assessed using the MyotonPRO device. Intraclass correlation coefficient (ICC) and absolute reliability was calculated and Bland-Altman analysis was performed to investigate the reliability. Result:The ICC of the intra-rater reliability was 0.76-0.90;Standard Errors of Measurement (SEM) was 0.05—0.09; SEM% was 4.63%—7.11%; Minimal Detectable Change (MDC) was 0.15—0.25; MDC% was 12.83%—19.70%. The ICC of the inter-rater reliability was 0.86—0.97;SEM was 0.03—0.08; SEM% was 3.42%—5.84%;MDC was 0.09—0.23; MDC% was 9.48%—16.19%. Bland-Altman plots showed that there was no systematic error. Conclusion:The MyotonPRO device has good reliability when it is used to measure the muscle elasticity of quadriceps femoris in healthy youth. It can also provide objective evidence for clinical work.
Keywords:MyotonPRO  quadriceps femoris  muscle elasticity  reliability